All after the
completion of the assembly, sensor calibration and guide rail parallel should
first, which guarantees within the 0.10MM. Then insert the grating digital
display meter corresponding to the jack, locking screws, open display box power,
starting the machine, observe whether the normal count. If not counting, should
carefully check the connection between the sensor and the digital display box if
the count can be moved to the end of the reading head shell, zero digital
display box, mobile reading head, digital display box display observation value
and the standard numerical control should be read, in the error range (range
2SM). If the ultra poor, carefully check the installation location and size to
meet the requirements.
Measuring range: 0mm ~ 3000mm,
Resolution is divided into: 1um (0.001mm) and 5um
(0.005mm)
One way repeatability accuracy: 2um (0.002mm), reverse
repetition accuracy 3um (0.003mm)
Pitch: 0.02mm gate lines: 50 lines of /mm, the
reaction rate: 120m/min (0.005mm) 20m/min (0.001mm)
Operating temperature: 0-45 DEG C, storage
temperature: -20 DEG C, -70 C, output signal: TTL, EIA-422-A
Supply voltage: DC 5V + 5% and 24V,
Cable length: 2.5m (longer)
Reading head rolling system: 0M-1M vertical type five
bearing rolling system, 1M-3M 45 degree five bearing rolling system
Optical measurement system: transmission infrared
light measurement system, infrared wavelength 880nm